Comparison of the BTE tech EVJ handgrip to the Jamar hand dynamometer for measuring grip strength
Journal article   Peer reviewed

Comparison of the BTE tech EVJ handgrip to the Jamar hand dynamometer for measuring grip strength

Edwin J. Myers, David S. Dominguez, Melissa E. Guigliano and Melissa G. Mallory
Technology and disability, Vol.33(3), pp.221-227
08-17-2021

Abstract

INTRODUCTION: This study compared the inter-instrument reliability of the Jamar Hand Dynamometer (Jamar) to the BTE EVJ Handgrip tool (EVJ). The EVJ is a new digital handgrip instrument designed by BTE Technologies Inc. (BTE Tech) utilizing Bluetooth technology to automatically upload data to a computer or electronic health record. METHOD: This repeated measure study engaged 338 participants (N= 199 females, N= 139 males) ages 20 to 50. All participants were tested on both devices following the protocol established by the American Society of Hand Therapists (ASHT). RESULTS: Comparative statistics included ICC values (0.81–0.84), Pearson R (correlation coefficient) (0.68–0.77), and Bland Altman plots (93–95% of data within 2 SD), indicating good inter-instrument agreement. CONCLUSIONS: The EVJ Handgrip, developed by BTE Technologies, demonstrated good inter-instrument reliability with the Jamar Hand Dynamometer and may be reliable to use when referencing the published normative values in the clinical environment.